Adam Kerr
Co-Director CNI / Sr Research Engineer
Electrical Engineering
Administrative Appointments
-
Research Director, Stanford Center for Cognitive and Neurobiological Imaging (2017 - Present)
All Publications
-
Spherical echo-planar time-resolved imaging (sEPTI) for rapid 3D quantitative T 2 * and susceptibility imaging.
Magnetic resonance in medicine
2024
Abstract
To develop a 3D spherical EPTI (sEPTI) acquisition and a comprehensive reconstruction pipeline for rapid high-quality whole-brain submillimeter T 2 * $$ {\mathrm{T}}_2^{\ast } $$ and QSM quantification.For the sEPTI acquisition, spherical k-space coverage is utilized with variable echo-spacing and maximum kx ramp-sampling to improve efficiency and signal incoherency compared to existing EPTI approaches. For reconstruction, an iterative rank-shrinking B0 estimation and odd-even high-order phase correction algorithms were incorporated into the reconstruction to better mitigate artifacts from field imperfections. A physics-informed unrolled network was utilized to boost the SNR, where 1-mm and 0.75-mm isotropic whole-brain imaging were performed in 45 and 90 s at 3 T, respectively. These protocols were validated through simulations, phantom, and in vivo experiments. Ten healthy subjects were recruited to provide sufficient data for the unrolled network. The entire pipeline was validated on additional five healthy subjects where different EPTI sampling approaches were compared. Two additional pediatric patients with epilepsy were recruited to demonstrate the generalizability of the unrolled reconstruction.sEPTI achieved 1.4 × $$ \times $$ faster imaging with improved image quality and quantitative map precision compared to existing EPTI approaches. The B0 update and the phase correction provide improved reconstruction performance with lower artifacts. The unrolled network boosted the SNR, achieving high-quality T 2 * $$ {\mathrm{T}}_2^{\ast } $$ and QSM quantification with single average data. High-quality reconstruction was also obtained in the pediatric patients using this network.sEPTI achieved whole-brain distortion-free multi-echo imaging and T 2 * $$ {\mathrm{T}}_2^{\ast } $$ and QSM quantification at 0.75 mm in 90 s which has the potential to be useful for wide clinical applications.
View details for DOI 10.1002/mrm.30255
View details for PubMedID 39250435
-
Rapid and accurate navigators for motion and B0 tracking using QUEEN: Quantitatively enhanced parameter estimation from navigators.
Magnetic resonance in medicine
2024
Abstract
To develop a framework that jointly estimates rigid motion and polarizing magnetic field (B0 ) perturbations ( δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ ) for brain MRI using a single navigator of a few milliseconds in duration, and to additionally allow for navigator acquisition at arbitrary timings within any type of sequence to obtain high-temporal resolution estimates.Methods exist that match navigator data to a low-resolution single-contrast image (scout) to estimate either motion or δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ . In this work, called QUEEN (QUantitatively Enhanced parameter Estimation from Navigators), we propose combined motion and δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ estimation from a fast, tailored trajectory with arbitrary-contrast navigator data. To this end, the concept of a quantitative scout (Q-Scout) acquisition is proposed from which contrast-matched scout data is predicted for each navigator. Finally, navigator trajectories, contrast-matched scout, and δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ are integrated into a motion-informed parallel-imaging framework.Simulations and in vivo experiments show the need to model δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ to obtain accurate motion parameters estimated in the presence of strong δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ . Simulations confirm that tailored navigator trajectories are needed to robustly estimate both motion and δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ . Furthermore, experiments show that a contrast-matched scout is needed for parameter estimation from multicontrast navigator data. A retrospective, in vivo reconstruction experiment shows improved image quality when using the proposed Q-Scout and QUEEN estimation.We developed a framework to jointly estimate rigid motion parameters and δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ from navigators. Combing a contrast-matched scout with the proposed trajectory allows for navigator deployment in almost any sequence and/or timing, which allows for higher temporal-resolution motion and δ B 0 $$ \delta {\mathbf{B}}_{\mathbf{0}} $$ estimates.
View details for DOI 10.1002/mrm.29976
View details for PubMedID 38173304
-
High-resolution myelin-water fraction and quantitative relaxation mapping using 3D ViSTa-MR fingerprinting.
Magnetic resonance in medicine
2023
Abstract
This study aims to develop a high-resolution whole-brain multi-parametric quantitative MRI approach for simultaneous mapping of myelin-water fraction (MWF), T1 , T2 , and proton-density (PD), all within a clinically feasible scan time.We developed 3D visualization of short transverse relaxation time component (ViSTa)-MRF, which combined ViSTa technique with MR fingerprinting (MRF), to achieve high-fidelity whole-brain MWF and T1 /T2 /PD mapping on a clinical 3T scanner. To achieve fast acquisition and memory-efficient reconstruction, the ViSTa-MRF sequence leverages an optimized 3D tiny-golden-angle-shuffling spiral-projection acquisition and joint spatial-temporal subspace reconstruction with optimized preconditioning algorithm. With the proposed ViSTa-MRF approach, high-fidelity direct MWF mapping was achieved without a need for multicompartment fitting that could introduce bias and/or noise from additional assumptions or priors.The in vivo results demonstrate the effectiveness of the proposed acquisition and reconstruction framework to provide fast multi-parametric mapping with high SNR and good quality. The in vivo results of 1 mm- and 0.66 mm-isotropic resolution datasets indicate that the MWF values measured by the proposed method are consistent with standard ViSTa results that are 30× slower with lower SNR. Furthermore, we applied the proposed method to enable 5-min whole-brain 1 mm-iso assessment of MWF and T1 /T2 /PD mappings for infant brain development and for post-mortem brain samples.In this work, we have developed a 3D ViSTa-MRF technique that enables the acquisition of whole-brain MWF, quantitative T1 , T2 , and PD maps at 1 and 0.66 mm isotropic resolution in 5 and 15 min, respectively. This advancement allows for quantitative investigations of myelination changes in the brain.
View details for DOI 10.1002/mrm.29990
View details for PubMedID 38156945
-
High-resolution myelin-water fraction and quantitative relaxation mapping using 3D ViSTa-MR fingerprinting.
ArXiv
2023
Abstract
This study aims to develop a high-resolution whole-brain multi-parametric quantitative MRI approach for simultaneous mapping of myelin-water fraction (MWF), T1, T2, and proton-density (PD), all within a clinically feasible scan time.We developed 3D ViSTa-MRF, which combined Visualization of Short Transverse relaxation time component (ViSTa) technique with MR Fingerprinting (MRF), to achieve high-fidelity whole-brain MWF and T1/T2/PD mapping on a clinical 3T scanner. To achieve fast acquisition and memory-efficient reconstruction, the ViSTa-MRF sequence leverages an optimized 3D tiny-golden-angle-shuffling spiral-projection acquisition and joint spatial-temporal subspace reconstruction with optimized preconditioning algorithm. With the proposed ViSTa-MRF approach, high-fidelity direct MWF mapping was achieved without a need for multi-compartment fitting that could introduce bias and/or noise from additional assumptions or priors.The in-vivo results demonstrate the effectiveness of the proposed acquisition and reconstruction framework to provide fast multi-parametric mapping with high SNR and good quality. The in-vivo results of 1mm- and 0.66mm-iso datasets indicate that the MWF values measured by the proposed method are consistent with standard ViSTa results that are 30x slower with lower SNR. Furthermore, we applied the proposed method to enable 5-minute whole-brain 1mm-iso assessment of MWF and T1/T2/PD mappings for infant brain development and for post-mortem brain samples.In this work, we have developed a 3D ViSTa-MRF technique that enables the acquisition of whole-brain MWF, quantitative T1, T2, and PD maps at 1mm and 0.66mm isotropic resolution in 5 and 15 minutes, respectively. This advancement allows for quantitative investigations of myelination changes in the brain.
View details for DOI 10.1016/j.neuroimage.2018.04.017
View details for PubMedID 38196746
View details for PubMedCentralID PMC10775347
-
DTI-MR fingerprinting for rapid high-resolution whole-brain T1 , T2 , proton density, ADC, and fractional anisotropy mapping.
Magnetic resonance in medicine
2023
Abstract
This study aims to develop a high-efficiency and high-resolution 3D imaging approach for simultaneous mapping of multiple key tissue parameters for routine brain imaging, including T1 , T2 , proton density (PD), ADC, and fractional anisotropy (FA). The proposed method is intended for pushing routine clinical brain imaging from weighted imaging to quantitative imaging and can also be particularly useful for diffusion-relaxometry studies, which typically suffer from lengthy acquisition time.To address challenges associated with diffusion weighting, such as shot-to-shot phase variation and low SNR, we integrated several innovative data acquisition and reconstruction techniques. Specifically, we used M1-compensated diffusion gradients, cardiac gating, and navigators to mitigate phase variations caused by cardiac motion. We also introduced a data-driven pre-pulse gradient to cancel out eddy currents induced by diffusion gradients. Additionally, to enhance image quality within a limited acquisition time, we proposed a data-sharing joint reconstruction approach coupled with a corresponding sequence design.The phantom and in vivo studies indicated that the T1 and T2 values measured by the proposed method are consistent with a conventional MR fingerprinting sequence and the diffusion results (including diffusivity, ADC, and FA) are consistent with the spin-echo EPI DWI sequence.The proposed method can achieve whole-brain T1 , T2 , diffusivity, ADC, and FA maps at 1-mm isotropic resolution within 10 min, providing a powerful tool for investigating the microstructural properties of brain tissue, with potential applications in clinical and research settings.
View details for DOI 10.1002/mrm.29916
View details for PubMedID 37936313
-
High-fidelity mesoscale in-vivo diffusion MRI through gSlider-BUDA and circular EPI with S-LORAKS reconstruction.
NeuroImage
2023: 120168
Abstract
To develop a high-fidelity diffusion MRI acquisition and reconstruction framework with reduced echo-train-length for less T2* image blurring compared to typical highly accelerated echo-planar imaging (EPI) acquisitions at sub-millimeter isotropic resolution.We first proposed a circular-EPI trajectory with partial Fourier sampling on both the readout and phase-encoding directions to minimize the echo-train-length and echo time. We then utilized this trajectory in an interleaved two-shot EPI acquisition with reversed phase-encoding polarity, to aid in the correction of off-resonance-induced image distortions and provide complementary k-space coverage in the missing partial Fourier regions. Using model-based reconstruction with structured low-rank constraint and smooth phase prior, we corrected the shot-to-shot phase variations across the two shots and recover the missing k-space data. Finally, we combined the proposed acquisition/reconstruction framework with an SNR-efficient RF-encoded simultaneous multi-slab technique, termed gSlider, to achieve high-fidelity 720μm and 500μm isotropic resolution in-vivo diffusion MRI.Both simulation and in-vivo results demonstrate the effectiveness of the proposed acquisition and reconstruction framework to provide distortion-corrected diffusion imaging at the mesoscale with markedly reduced T2*-blurring. The in-vivo results of 720μm and 500μm datasets show high-fidelity diffusion images with reduced image blurring and echo time using the proposed approaches.The proposed method provides high-quality distortion-corrected diffusion-weighted images with ∼40% reduction in the echo-train-length and T2* blurring at 500μm-isotropic-resolution compared to standard multi-shot EPI.
View details for DOI 10.1016/j.neuroimage.2023.120168
View details for PubMedID 37187364
-
Measuring brain beats: Cardiac-aligned fast functional magnetic resonance imaging signals.
Human brain mapping
2022
Abstract
Blood and cerebrospinal fluid (CSF) pulse and flow throughout the brain, driven by the cardiac cycle. These fluid dynamics, which are essential to healthy brain function, are characterized by several noninvasive magnetic resonance imaging (MRI) methods. Recent developments in fast MRI, specifically simultaneous multislice acquisition methods, provide a new opportunity to rapidly and broadly assess cardiac-driven flow, including CSF spaces, surface vessels and parenchymal vessels. We use these techniques to assess blood and CSF flow dynamics in brief (3.5min) scans on a conventional 3T MRI scanner in five subjects. Cardiac pulses are measured with a photoplethysmography (PPG) on the index finger, along with functional MRI (fMRI) signals in the brain. We, retrospectively, align the fMRI signals to the heartbeat. Highly reliable cardiac-gated fMRI temporal signals are observed in CSF and blood on the timescale of one heartbeat (test-retest reliability within subjects R2 >50%). In blood vessels, a local minimum is observed following systole. In CSF spaces, the ventricles and subarachnoid spaces have a local maximum following systole instead. Slower resting-state scans with slice timing, retrospectively, aligned to the cardiac pulse, reveal similar cardiac-gated responses. The cardiac-gated measurements estimate the amplitude and phase of fMRI pulsations in the CSF relative to those in the arteries, an estimate of the local intracranial impedance. Cardiac aligned fMRI signals can provide new insights about fluid dynamics or diagnostics for diseases where these dynamics are important.
View details for DOI 10.1002/hbm.26128
View details for PubMedID 36308417
-
Convergence, preliminary findings and future directions across the four human connectome projects investigating mood and anxiety disorders.
NeuroImage
2021: 118694
Abstract
In this paper we provide an overview of the rationale, methods, and preliminary results of the four Connectome Studies Related to Human Disease investigating mood and anxiety disorders. The first study, "Dimensional connectomics of anxious misery" (HCP-DAM), characterizes brain-symptom relations of a transdiagnostic sample of anxious misery disorders. The second study, "Human connectome Project for disordered emotional states" (HCP-DES), tests a hypothesis-driven model of brain circuit dysfunction in a sample of untreated young adults with symptoms of depression and anxiety. The third study, "Perturbation of the treatment resistant depression connectome by fast-acting therapies" (HCP-MDD), quantifies alterations of the structural and functional connectome as a result of three fast-acting interventions: electroconvulsive therapy, serial ketamine therapy, and total sleep deprivation. Finally, the fourth study, "Connectomes related to anxiety and depression in adolescents" (HCP-ADA), investigates developmental trajectories of subtypes of anxiety and depression in adolescence. The four projects use comparable and standardized Human Connectome Project magnetic resonance imaging (MRI) protocols, including structural MRI, diffusion-weighted MRI, and both task and resting state functional MRI. All four projects also conducted comprehensive and convergent clinical and neuropsychological assessments, including (but not limited to) demographic information, clinical diagnoses, symptoms of mood and anxiety disorders, negative and positive affect, cognitive function, and exposure to early life stress. The first round of analyses conducted in the four projects offered novel methods to investigate relations between functional connectomes and self-reports in large datasets, identified new functional correlates of symptoms of mood and anxiety disorders, characterized the trajectory of connectome-symptom profiles over time, and quantified the impact of novel treatments on aberrant connectivity. Taken together, the data obtained and reported by the four Connectome Studies Related to Human Disease investigating mood and anxiety disorders describe a rich constellation of convergent biological, clinical, and behavioral phenotypes that span the peak ages for the onset of emotional disorders. These data are being prepared for open sharing with the scientific community following screens for quality by the Connectome Coordinating Facility (CCF). The CCF also plans to release data from all projects that have been pre-processed using identical state-of-the-art pipelines. The resultant dataset will give researchers the opportunity to pool complementary data across the four projects to study circuit dysfunctions that may underlie mood and anxiety disorders, to map cohesive relations among circuits and symptoms, and to probe how these relations change as a function of age and acute interventions. This large and combined dataset may also be ideal for using data-driven analytic approaches to inform neurobiological targets for future clinical trials and interventions focused on clinical or behavioral outcomes.
View details for DOI 10.1016/j.neuroimage.2021.118694
View details for PubMedID 34732328
-
Utilizing the Wavelet Transform's Structure in Compressed Sensing.
Signal, image and video processing
2021; 15 (7): 1407-1414
Abstract
Compressed sensing has empowered quality image reconstruction with fewer data samples than previously thought possible. These techniques rely on a sparsifying linear transformation. The Daubechies wavelet transform is commonly used for this purpose. In this work, we take advantage of the structure of this wavelet transform and identify an affine transformation that increases the sparsity of the result. After inclusion of this affine transformation, we modify the resulting optimization problem to comply with the form of the Basis Pursuit Denoising problem. Finally, we show theoretically that this yields a lower bound on the error of the reconstruction and present results where solving this modified problem yields images of higher quality for the same sampling patterns using both magnetic resonance and optical images.
View details for DOI 10.1007/s11760-021-01872-y
View details for PubMedID 34531930
View details for PubMedCentralID PMC8439112
-
Frequency Drift in MR Spectroscopy at 3T.
NeuroImage
2021: 118430
Abstract
PURPOSE: Heating of gradient coils and passive shim components is a common cause of instability in the B0 field, especially when gradient intensive sequences are used. The aim of the study was to set a benchmark for typical drift encountered during MR spectroscopy (MRS) to assess the need for real-time field-frequency locking on MRI scanners by comparing field drift data from a large number of sites.METHOD: A standardized protocol was developed for 80 participating sites using 99 3T MR scanners from 3 major vendors. Phantom water signals were acquired before and after an EPI sequence. The protocol consisted of: minimal preparatory imaging; a short pre-fMRI PRESS; a ten-minute fMRI acquisition; and a long post-fMRI PRESS acquisition. Both pre- and post-fMRI PRESS were non-water suppressed. Real-time frequency stabilization/adjustment was switched off when appropriate. Sixty scanners repeated the protocol for a second dataset. In addition, a three-hour post-fMRI MRS acquisition was performed at one site to observe change of gradient temperature and drift rate. Spectral analysis was performed using MATLAB. Frequency drift in pre-fMRI PRESS data were compared with the first 5:20 minutes and the full 30:00 minutes of data after fMRI. Median (interquartile range) drifts were measured and showed in violin plot. Paired t-tests were performed to compare frequency drift pre- and post-fMRI. A simulated in vivo spectrum was generated using FID-A to visualize the effect of the observed frequency drifts. The simulated spectrum was convolved with the frequency trace for the most extreme cases. Impacts of frequency drifts on NAA and GABA were also simulated as a function of linear drift. Data from the repeated protocol were compared with the corresponding first dataset using Pearson's and intraclass correlation coefficients (ICC).RESULTS: Of the data collected from 99 scanners, 4 were excluded due to various reasons. Thus, data from 95 scanners were ultimately analyzed. For the first 5:20 min (64 transients), median (interquartile range) drift was 0.44 (1.29) Hz before fMRI and 0.83 (1.29) Hz after. This increased to 3.15 (4.02) Hz for the full 30 min (360 transients) run. Average drift rates were 0.29 Hz/min before fMRI and 0.43 Hz/min after. Paired t-tests indicated that drift increased after fMRI, as expected (p < 0.05). Simulated spectra convolved with the frequency drift showed that the intensity of the NAA singlet was reduced by up to 26%, 44 % and 18% for GE, Philips and Siemens scanners after fMRI, respectively. ICCs indicated good agreement between datasets acquired on separate days. The single site long acquisition showed drift rate was reduced to 0.03 Hz/min approximately three hours after fMRI.DISCUSSION: This study analyzed frequency drift data from 95 3T MRI scanners. Median levels of drift were relatively low (5-min average under 1 Hz), but the most extreme cases suffered from higher levels of drift. The extent of drift varied across scanners which both linear and nonlinear drifts were observed.
View details for DOI 10.1016/j.neuroimage.2021.118430
View details for PubMedID 34314848
-
The human connectome project for disordered emotional states: Protocol and rationale for a research domain criteria study of brain connectivity in young adult anxiety and depression.
NeuroImage
2020: 116715
Abstract
Through the Human Connectome Project (HCP) our understanding of the functional connectome of the healthy brain has been dramatically accelerated. Given the pressing public health need, we must increase our understanding of how connectome dysfunctions give rise to disordered mental states. Mental disorders arising from high levels of negative emotion or from the loss of positive emotional experience affect over 400 million people globally. Such states of disordered emotion cut across multiple diagnostic categories of mood and anxiety disorders and are compounded by accompanying disruptions in cognitive function. Not surprisingly, these forms of psychopathology are the leading cause of disability worldwide. The Research Domain Criteria (RDoC) initiative spearheaded by NIMH offers a framework for characterizing the relations among connectome dysfunctions, anchored in neural circuits and phenotypic profiles of behavior and self-reported symptoms. Here, we report on our Connectomes Related to Human Disease protocol for integrating an RDoC framework with HCP protocols to characterize connectome dysfunctions in disordered emotional states, and present quality control data from a representative sample of participants. We focus on three RDoC domains and constructs most relevant to depression and anxiety: 1) loss and acute threat within the Negative Valence System (NVS) domain; 2) reward valuation and responsiveness within the Positive Valence System (PVS) domain; and 3) working memory and cognitive control within the Cognitive System (CS) domain. For 29 healthy controls, we present preliminary imaging data: functional magnetic resonance imaging collected in the resting state and in tasks matching our constructs of interest ("Emotion", "Gambling" and "Continuous Performance" tasks), as well as diffusion-weighted imaging. All functional scans demonstrated good signal-to-noise ratio. Established neural networks were robustly identified in the resting state condition by independent component analysis. Processing of negative emotional faces significantly activated the bilateral dorsolateral prefrontal and occipital cortices, fusiform gyrus and amygdalae. Reward elicited a response in the bilateral dorsolateral prefrontal, parietal and occipital cortices, and in the striatum. Working memory was associated with activation in the dorsolateral prefrontal, parietal, motor, temporal and insular cortices, in the striatum and cerebellum. Diffusion tractography showed consistent profiles of fractional anisotropy along known white matter tracts. We also show that results are comparable to those in a matched sample from the HCP Healthy Young Adult data release. These preliminary data provide the foundation for acquisition of 250 subjects who are experiencing disordered emotional states. When complete, these data will be used to develop a neurobiological model that maps connectome dysfunctions to specific behaviors and symptoms.
View details for DOI 10.1016/j.neuroimage.2020.116715
View details for PubMedID 32147367
-
Combined T2 -preparation and multidimensional outer volume suppression for coronary artery imaging with 3D cones trajectories.
Magnetic resonance in medicine
2019
Abstract
PURPOSE: To develop a modular magnetization preparation sequence for combined T2 -preparation and multidimensional outer volume suppression (OVS) for coronary artery imaging.METHODS: A combined T2 -prepared 1D OVS sequence with fat saturation was defined to contain a 90°-60 180°60 composite nonselective tip-down pulse, two 180°Y hard pulses for refocusing, and a -90° spectral-spatial sinc tip-up pulse. For 2D OVS, 2 modules were concatenated, selective in X and then Y. Bloch simulations predicted robustness of the sequence to B0 and B1 inhomogeneities. The proposed sequence was compared with a T2 -prepared 2D OVS sequence proposed by Luo et al, which uses a spatially selective 2D spiral tip-up. The 2 sequences were compared in phantom studies and in vivo coronary artery imaging studies with a 3D cones trajectory.RESULTS: Phantom results demonstrated superior OVS for the proposed sequence compared with the Luo sequence. In studies on 15 healthy volunteers, the proposed sequence had superior image edge profile acutance values compared with the Luo sequence for the right (P < .05) and left (P < .05) coronary arteries, suggesting superior vessel sharpness. The proposed sequence also had superior signal-to-noise ratio (P < .05) and passband-to-stopband ratio (P < .05). Reader scores and reader preference indicated superior coronary image quality of the proposed sequence for both the right (P < .05) and left (P < .05) coronary arteries.CONCLUSION: The proposed sequence with concatenated 1D spatially selective tip-ups and integrated fat saturation has superior image quality and suppression compared with the Luo sequence with 2D spatially selective tip-up.
View details for DOI 10.1002/mrm.28057
View details for PubMedID 31691350
-
SMS MUSSELS: A navigator-free reconstruction for simultaneous multi-slice-accelerated multi-shot diffusion weighted imaging.
Magnetic resonance in medicine
2019
Abstract
PURPOSE: To introduce a novel reconstruction method for simultaneous multi-slice (SMS)-accelerated multi-shot diffusion weighted imaging (ms-DWI).METHODS: SMS acceleration using blipped-CAIPI schemes have been proposed to speed up the acquisition of ms-DWIs. The reconstruction of the data requires (a) phase compensation to combine data from different shots and (b) slice unfolding to separate the data of different slices. The traditional approaches first estimate the phase maps corresponding to each shot and slice which are then employed to iteratively recover the slice unfolded DWIs without phase artifacts. In contrast, the proposed reconstruction directly recovers the slice-unfolded k-space data of the multiple shots for each slice in a single-step recovery scheme. The proposed method is enabled by the low-rank property inherent in the k-space samples of ms-DW acquisition. This enabled to formulate a joint recovery scheme that simultaneously (a) unfolds the k-space data of each slice using a SENSE-based scheme and (b) recover the missing k-space samples in each slice of the multi-shot acquisition employing a structured low-rank matrix completion. Additional smoothness regularization is also utilized for higher acceleration factors. The proposed joint recovery is tested on simulated and in vivo data and compared to similar un-navigated methods.RESULTS: Our experiments show effective slice unfolding and successful recovery of DWIs with minimal phase artifacts using the proposed method. The performance is comparable to existing methods at low acceleration factors and better than existing methods for higher acceleration factors.CONCLUSIONS: For the slice accelerations considered in this study, the proposed method can successfully recover DWIs from SMS-accelerated ms-DWI acquisitions.
View details for DOI 10.1002/mrm.27924
View details for PubMedID 31403223
-
MR susceptibility contrast imaging using a 2D simultaneous multi-slice gradient-echo sequence at 7T.
PloS one
2019; 14 (7): e0219705
Abstract
PURPOSE: To develop a 7T simultaneous multi-slice (SMS) 2D gradient-echo sequence for susceptibility contrast imaging, and to compare its quality to 3D imaging.METHODS: A frequency modulated and phase cycled RF pulse was designed to simultaneously excite multiple slices in multi-echo 2D gradient-echo imaging. The imaging parameters were chosen to generate images with susceptibility contrast, including T2*-weighted magnitude/phase images, susceptibility-weighted images and quantitative susceptibility/R2* maps. To compare their image quality with 3D gradient-echo imaging, both 2D and 3D imaging were performed on 11 healthy volunteers and 4 patients with multiple sclerosis (MS). The signal to noise ratio (SNR) in gray and white matter and their contrast to noise ratio (CNR) was simulated for the 2D and 3D magnitude images using parameters from the imaging. The experimental SNRs and CNRs were measured in gray/white matter and deep gray matter structures on magnitude, phase, R2* and QSM images from volunteers and the visibility of MS lesions on these images from patients was visually rated. All SNRs and CNRs were compared between the 2D and 3D imaging using a paired t-test.RESULTS: Although the 3D magnitude images still had significantly higher SNRs (by 13.0~17.6%), the 2D magnitude and QSM images generated significantly higher gray/white matter or globus pallidus/putamen contrast (by 13.3~87.5%) and significantly higher MS lesion contrast (by 5.9~17.3%).CONCLUSION: 2D SMS gradient-echo imaging can serve as an alternative to often used 3D imaging to obtain susceptibility-contrast-weighted images, with an advantage of providing better image contrast and MS lesion sensitivity.
View details for DOI 10.1371/journal.pone.0219705
View details for PubMedID 31314813
-
Technique development of 3D dynamic CS-EPSI for hyperpolarized C-13 pyruvate MR molecular imaging of human prostate cancer
MAGNETIC RESONANCE IN MEDICINE
2018; 80 (5): 2062–72
Abstract
The purpose of this study was to develop a new 3D dynamic carbon-13 compressed sensing echoplanar spectroscopic imaging (EPSI) MR sequence and test it in phantoms, animal models, and then in prostate cancer patients to image the metabolic conversion of hyperpolarized [1-13 C]pyruvate to [1-13 C]lactate with whole gland coverage at high spatial and temporal resolution.A 3D dynamic compressed sensing (CS)-EPSI sequence with spectral-spatial excitation was designed to meet the required spatial coverage, time and spatial resolution, and RF limitations of the 3T MR scanner for its clinical translation for prostate cancer patient imaging. After phantom testing, animal studies were performed in rats and transgenic mice with prostate cancers. For patient studies, a GE SPINlab polarizer (GE Healthcare, Waukesha, WI) was used to produce hyperpolarized sterile GMP [1-13 C]pyruvate. 3D dynamic 13 C CS-EPSI data were acquired starting 5 s after injection throughout the gland with a spatial resolution of 0.5 cm3 , 18 time frames, 2-s temporal resolution, and 36 s total acquisition time.Through preclinical testing, the 3D CS-EPSI sequence developed in this project was shown to provide the desired spectral, temporal, and spatial 5D HP 13 C MR data. In human studies, the 3D dynamic HP CS-EPSI approach provided first-ever simultaneously volumetric and dynamic images of the LDH-catalyzed conversion of [1-13 C]pyruvate to [1-13 C]lactate in a biopsy-proven prostate cancer patient with full gland coverage.The results demonstrate the feasibility to characterize prostate cancer metabolism in animals, and now patients using this new 3D dynamic HP MR technique to measure kPL , the kinetic rate constant of [1-13 C]pyruvate to [1-13 C]lactate conversion.
View details for PubMedID 29575178
View details for PubMedCentralID PMC6107425
-
Advantages of Short Repetition Time Resting-State Functional MRI Enabled by Simultaneous Multi-slice Imaging.
Journal of neuroscience methods
2018
Abstract
BACKGROUND: Recent advancements in simultaneous multi-slice (SMS) imaging techniques have enabled whole-brain resting-state fMRI (rs-fMRI) scanning at sub-second temporal resolution, providing spectral ranges much wider than the typically used range of 0.01-0.1Hz. However, the advantages of this accelerated acquisition for rs-fMRI have not been evaluated.NEW METHOD: In this study, we used SMS Echo Planar Imaging (EPI) to probe whole-brain functional connectivity with a short repetition time (TR=350ms) and compared it with standard EPI with a longer TR of 2000ms. We determined the effect of scan length and investigated the temporal filtration strategies that optimize results based on metrics of signal-noise separation and test-retest reliability using both seed-based and independent component analysis (ICA).RESULTS: We found that use of either the entire frequency range of 0.01-1.4Hz or the entire frequency range with the exclusion of typical cardiac and respiratory frequency values tended to provide the best functional connectivity maps.COMPARISON WITH EXISTING METHODS: We found that the SMS-acquired rs-fMRI scans had improved the signal-noise separation, while preserving the same level of test-retest reliability compared to conventional EPI, and enabled the detection of reliable functional connectivity networks with scan times as short as 3minutes.CONCLUSIONS: Our findings suggest that whole-brain rs-fMRI studies may benefit from the increased temporal resolution enabled by the SMS-EPI acquisition, leading to drastic scan time reductions, which in turn should enable the more widespread use of rs-fMRI in clinical research protocols.
View details for PubMedID 30300699
-
Body diffusion-weighted imaging using magnetization prepared single-shot fast spin echo and extended parallel imaging signal averaging
MAGNETIC RESONANCE IN MEDICINE
2018; 79 (6): 3032–44
Abstract
This work demonstrates a magnetization prepared diffusion-weighted single-shot fast spin echo (SS-FSE) pulse sequence for the application of body imaging to improve robustness to geometric distortion. This work also proposes a scan averaging technique that is superior to magnitude averaging and is not subject to artifacts due to object phase.This single-shot sequence is robust against violation of the Carr-Purcell-Meiboom-Gill (CPMG) condition. This is achieved by dephasing the signal after diffusion weighting and tipping the MG component of the signal onto the longitudinal axis while the non-MG component is spoiled. The MG signal component is then excited and captured using a traditional SS-FSE sequence, although the echo needs to be recalled prior to each echo. Extended Parallel Imaging (ExtPI) averaging is used where coil sensitivities from the multiple acquisitions are concatenated into one large parallel imaging (PI) problem. The size of the PI problem is reduced by SVD-based coil compression which also provides background noise suppression. This sequence and reconstruction are evaluated in simulation, phantom scans, and in vivo abdominal clinical cases.Simulations show that the sequence generates a stable signal throughout the echo train which leads to good image quality. This sequence is inherently low-SNR, but much of the SNR can be regained through scan averaging and the proposed ExtPI reconstruction. In vivo results show that the proposed method is able to provide diffusion encoded images while mitigating geometric distortion artifacts compared to EPI.This work presents a diffusion-prepared SS-FSE sequence that is robust against the violation of the CPMG condition while providing diffusion contrast in clinical cases. Magn Reson Med 79:3032-3044, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
View details for PubMedID 29044721
-
Slice profile effects on nCPMG SS-FSE.
Magnetic resonance in medicine
2017
Abstract
To determine the effects of the RF refocusing pulse profile on the magnitude of the transverse signal smoothness throughout the echo train in non-Carr-Purcell-Meiboom-Gill (nCPMG) single-shot fast spin echo (SS-FSE) imaging and to design an RF refocusing pulse that provides improved signal stability. THEORY AND METHODS: nCPMG SS-FSE quadratic phase modulation requires sufficiently high and uniform refocusing flip angle to achieve a stable signal. Typically, refocusing pulses used in SS-FSE sequences are designed for minimum duration to minimize echo spacing and as a consequence have poor selectivity. However, delay-insensitive variable rate excitation Shinnar-Le Roux (DV-SLR) refocusing pulses can achieve both improved selectivity as well as a short duration. This class of RF pulse is compared against a traditional low time-bandwidth refocusing pulse in a nCPMG SS-FSE in simulation, phantom, and in vivo.DV-SLR pulses achieve a more stable signal in simulation, phantom, and in vivo cases while maintaining an appropriately short duration as well as not dramatically increasing specific absorption rate (SAR) accumulation.The nCPMG SS-FSE method demonstrates improved robustness when a more selective refocusing pulse is used. Refocusing pulses that use a time-varying excitation gradient can achieve this selectivity while maintaining short echo spacing. Magn Reson Med, 2017. © 2017 International Society for Magnetic Resonance in Medicine.
View details for DOI 10.1002/mrm.26694
View details for PubMedID 28370409
-
Body Diffusion Weighted Imaging Using Non-CPMG Fast Spin Echo
IEEE TRANSACTIONS ON MEDICAL IMAGING
2017; 36 (2): 549-559
Abstract
SS-FSE is a fast technique that does not suffer from off-resonance distortions to the degree that EPI does. Unlike EPI, SS-FSE is ill-suited to diffusion weighted imaging (DWI) due to the Carr-Purcell-Meiboom-Geill (CPMG) condition. Non- CPMG phase cycling does accommodate SS-FSE and DWI but places constraints on reconstruction, which are resolved here through parallel imaging. Additionally, improved echo stability can be achieved by using short duration and highly selective DIVERSE radiofrequency pulses. Here, signal-to-noise ratio (SNR) comparisons between EPI and nCPMG SS-FSE acquisitions and reconstruction techniques give similar values. Diffusion imaging with nCPMG SS-FSE gives similar SNR to an EPI acquisition, though apparent diffusion coefficient values are higher than seen with EPI. In vivo images have good image quality with little distortion. This method has the ability to capture distortionfree DWI images near areas of significant off-resonance as well as preserve adequate SNR. Parallel imaging and DIVERSE refocusing RF pulses allow shorter ETL compared to previous implementations and thus reduces phase encode direction blur and SAR accumulation.
View details for DOI 10.1109/TMI.2016.2622238
View details for Web of Science ID 000396115800019
-
Measuring B-1 distributions by B-1 phase encoding
MAGNETIC RESONANCE IN MEDICINE
2017; 77 (1): 229-236
Abstract
We propose a method to acquire B1 distribution plots by encoding in B1 instead of image space. Using this method, B1 data is acquired in a different way from traditional spatial B1 mapping, and allows for quick measurement of high dynamic range B1 data.To encode in B1, we acquire multiple projections of a slice, each along the same direction, but using a different phase sensitivity to B1. Using a convex optimization formulation, we reconstruct histograms of the B1 distribution estimates of the slice.We verify in vivo B1 distribution measurements by comparing measured distributions to distributions calculated from reference spatial B1 maps using the Earth Mover's Distance. Phantom measurements using a surface coil show that for increased spatial B1 variations, measured B1 distributions using the proposed method more accurately estimate the distribution than a low-resolution spatial B1 map, resulting in a 37% Earth Mover's Distance decrease while using fewer measurements.We propose and validate the performance of a method to acquire B1 distribution information directly without acquiring a spatial B1 map. The method may provide faster estimates of a B1 field for applications that do not require spatial B1 localization, such as the transmit gain calibration of the scanner, particularly for high dynamic B1 ranges. Magn Reson Med 77:229-236, 2017. © 2016 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.26114
View details for Web of Science ID 000391038800024
View details for PubMedCentralID PMC4947573
-
Spectrally selective three-dimensional dynamic balanced steady-state free precession for hyperpolarized C-13 metabolic imaging with spectrally selective radiofrequency pulses.
Magnetic resonance in medicine
2016
Abstract
Balanced steady-state free precession (bSSFP) sequences can provide superior signal-to-noise ratio efficiency for hyperpolarized (HP) carbon-13 ((13) C) magnetic resonance imaging by efficiently utilizing the nonrecoverable magnetization, but managing their spectral response is challenging in the context of metabolic imaging. A new spectrally selective bSSFP sequence was developed for fast imaging of multiple HP (13) C metabolites with high spatiotemporal resolution.This novel approach for bSSFP spectral selectivity incorporates optimized short-duration spectrally selective radiofrequency pulses within a bSSFP pulse train and a carefully chosen repetition time to avoid banding artifacts.The sequence enabled subsecond 3D dynamic spectrally selective imaging of (13) C metabolites of copolarized [1-(13) C]pyruvate and [(13) C]urea at 2-mm isotropic resolution, with excellent spectral selectivity (∼100:1). The sequence was successfully tested in phantom studies and in vivo studies with normal mice.This sequence is expected to benefit applications requiring dynamic volumetric imaging of metabolically active (13) C compounds at high spatiotemporal resolution, including preclinical studies at high field and, potentially, clinical studies.Magn Reson Med, 2016. © 2016 International Society for Magnetic Resonance in Medicine.
View details for DOI 10.1002/mrm.26480
View details for PubMedID 27770458
-
Lowering the B1 threshold for improved BEAR B1 mapping.
Magnetic resonance in medicine
2016; 75 (3): 1262-1268
Abstract
Accurate measurement of the nonuniform transmit radiofrequency field is necessary for magnetic resonance imaging applications. The radiofrequency field excitation amplitude (B1 ) is often obtained by acquiring a B1 map. We modify the B1 estimation using adiabatic refocusing (BEAR) method to extend its range to lower B1 magnitudes.The BEAR method is a phase-based B1 mapping method, wherein hyperbolic secant pulses induce a phase sensitivity to B1 . The measurable B1 range is limited due to the adiabatic threshold of the pulses. We redesign the method to use flattened hyperbolic secant pulses, which have lower adiabatic thresholds. We optimize the flattened hyperbolic secant parameters to minimize phase sensitivity to frequency variations.We validate the performance of the new method via simulation and in vivo at 3T, and show that for n≤8, accurate B1 maps can be acquired using reduced nominal peak B1 values.The adiabatic threshold for the BEAR method is reduced with flattened hyperbolic secant pulses, which are optimized for accurate phase-to-B1 mapping over a frequency range, and allow for lower nominal B1 values. At 3T, the nominal B1 is decreased by 52% and the sensitivity to B1 is increased by a factor of 3.8. This can improve the method's applicability for measurement of low B1 . Magn Reson Med 75:1262-1268, 2016. © 2015 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.25711
View details for PubMedID 25846905
-
H-1 MR spectroscopic imaging of the prostate at 7T using spectral-spatial pulses
MAGNETIC RESONANCE IN MEDICINE
2016; 75 (3): 933–45
Abstract
To assess the feasibility of prostate (1)H MR spectroscopic imaging (MRSI) using low-power spectral-spatial (SPSP) pulses at 7T, exploiting accurate spectral selection and spatial selectivity simultaneously.A double spin-echo sequence was equipped with SPSP refocusing pulses with a spectral selectivity of 1 ppm. Three-dimensional prostate (1)H-MRSI at 7T was performed with the SPSP-MRSI sequence using an 8-channel transmit array coil and an endorectal receive coil in three patients with prostate cancer and in one healthy subject. No additional water or lipid suppression pulses were used.Prostate (1)H-MRSI could be obtained well within specific absorption rate (SAR) limits in a clinically feasible time (10 min). Next to the common citrate signals, the prostate spectra exhibited high spermine signals concealing creatine and sometimes also choline. Residual lipid signals were observed at the edges of the prostate because of limitations in spectral and spatial selectivity.It is possible to perform prostate (1)H-MRSI at 7T with a SPSP-MRSI sequence while using separate transmit and receive coils. This low-SAR MRSI concept provides the opportunity to increase spatial resolution of MRSI within reasonable scan times.
View details for PubMedID 25943445
-
Measuring B1 distributions by B1 phase encoding.
Magnetic resonance in medicine
2016
Abstract
We propose a method to acquire B1 distribution plots by encoding in B1 instead of image space. Using this method, B1 data is acquired in a different way from traditional spatial B1 mapping, and allows for quick measurement of high dynamic range B1 data.To encode in B1, we acquire multiple projections of a slice, each along the same direction, but using a different phase sensitivity to B1. Using a convex optimization formulation, we reconstruct histograms of the B1 distribution estimates of the slice.We verify in vivo B1 distribution measurements by comparing measured distributions to distributions calculated from reference spatial B1 maps using the Earth Mover's Distance. Phantom measurements using a surface coil show that for increased spatial B1 variations, measured B1 distributions using the proposed method more accurately estimate the distribution than a low-resolution spatial B1 map, resulting in a 37% Earth Mover's Distance decrease while using fewer measurements.We propose and validate the performance of a method to acquire B1 distribution information directly without acquiring a spatial B1 map. The method may provide faster estimates of a B1 field for applications that do not require spatial B1 localization, such as the transmit gain calibration of the scanner, particularly for high dynamic B1 ranges. Magn Reson Med 77:229-236, 2017. © 2016 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.26114
View details for PubMedID 26778689
View details for PubMedCentralID PMC4947573
-
Multiband RF pulses with improved performance via convex optimization
JOURNAL OF MAGNETIC RESONANCE
2016; 262: 81-90
Abstract
Selective RF pulses are commonly designed with the desired profile as a low pass filter frequency response. However, for many MRI and NMR applications, the spectrum is sparse with signals existing at a few discrete resonant frequencies. By specifying a multiband profile and releasing the constraint on "don't-care" regions, the RF pulse performance can be improved to enable a shorter duration, sharper transition, or lower peak B1 amplitude. In this project, a framework for designing multiband RF pulses with improved performance was developed based on the Shinnar-Le Roux (SLR) algorithm and convex optimization. It can create several types of RF pulses with multiband magnitude profiles, arbitrary phase profiles and generalized flip angles. The advantage of this framework with a convex optimization approach is the flexible trade-off of different pulse characteristics. Designs for specialized selective RF pulses for balanced SSFP hyperpolarized (HP) (13)C MRI, a dualband saturation RF pulse for (1)H MR spectroscopy, and a pre-saturation pulse for HP (13)C study were developed and tested.
View details for DOI 10.1016/j.jmr.2015.11.010
View details for Web of Science ID 000369877700013
View details for PubMedID 26754063
View details for PubMedCentralID PMC4716678
-
Controlling Radiofrequency-Induced Currents in Guidewires Using Parallel Transmit
MAGNETIC RESONANCE IN MEDICINE
2015; 74 (6): 1790-1802
Abstract
Elongated conductors, such as pacemaker leads, neurostimulator leads, and conductive guidewires used for interventional procedures can couple to the MRI radiofrequency (RF) transmit field, potentially causing dangerous tissue heating. The purpose of this study was to demonstrate the feasibility of using parallel transmit to control induced RF currents in elongated conductors, thereby reducing the RF heating hazard.Phantom experiments were performed on a four-channel parallel transmit system at 1.5T. Parallel transmit "null mode" excitations that induce minimal wire current were designed using coupling measurements derived from axial B1 (+) maps. The resulting current reduction performance was evaluated with B1 (+) maps, current sensor measurements, and fluoroptic temperature probe measurements.Null mode excitations reduced the maximum coupling mode current by factors ranging from 2 to 80. For the straight wire experiment, a current null imposed at a single wire location was sufficient to reduce tip heating below detectable levels. For longer insertion lengths and a curved geometry, imposing current nulls at two wire locations resulted in more distributed current reduction along the wire length.Parallel transmit can be used to create excitations that induce minimal RF current in elongated conductors, thereby decreasing the RF heating risk, while still allowing visualization of the surrounding volume.
View details for DOI 10.1002/mrm.25543
View details for Web of Science ID 000367737300031
View details for PubMedCentralID PMC4470871
-
Controlling radiofrequency-induced currents in guidewires using parallel transmit.
Magnetic resonance in medicine
2015; 74 (6): 1790-802
Abstract
Elongated conductors, such as pacemaker leads, neurostimulator leads, and conductive guidewires used for interventional procedures can couple to the MRI radiofrequency (RF) transmit field, potentially causing dangerous tissue heating. The purpose of this study was to demonstrate the feasibility of using parallel transmit to control induced RF currents in elongated conductors, thereby reducing the RF heating hazard.Phantom experiments were performed on a four-channel parallel transmit system at 1.5T. Parallel transmit "null mode" excitations that induce minimal wire current were designed using coupling measurements derived from axial B1 (+) maps. The resulting current reduction performance was evaluated with B1 (+) maps, current sensor measurements, and fluoroptic temperature probe measurements.Null mode excitations reduced the maximum coupling mode current by factors ranging from 2 to 80. For the straight wire experiment, a current null imposed at a single wire location was sufficient to reduce tip heating below detectable levels. For longer insertion lengths and a curved geometry, imposing current nulls at two wire locations resulted in more distributed current reduction along the wire length.Parallel transmit can be used to create excitations that induce minimal RF current in elongated conductors, thereby decreasing the RF heating risk, while still allowing visualization of the surrounding volume.
View details for DOI 10.1002/mrm.25543
View details for PubMedID 25521751
View details for PubMedCentralID PMC4470871
-
Chemical Shift Separation with Controlled Aliasing for Hyperpolarized C-13 Metabolic Imaging
MAGNETIC RESONANCE IN MEDICINE
2015; 74 (4): 978-989
Abstract
A chemical shift separation technique for hyperpolarized (13) C metabolic imaging with high spatial and temporal resolution was developed. Specifically, a fast three-dimensional pulse sequence and a reconstruction method were implemented to acquire signals from multiple (13) C species simultaneously with subsequent separation into individual images.A stack of flyback echo-planar imaging readouts and a set of multiband excitation radiofrequency pulses were designed to spatially modulate aliasing patterns of the acquired metabolite images, which translated the chemical shift separation problem into parallel imaging reconstruction problem. An eight-channel coil array was used for data acquisition and a parallel imaging method based on nonlinear inversion was developed to separate the aliased images.Simultaneous acquisitions of pyruvate and lactate in a phantom study and in vivo rat experiments were performed. The results demonstrated successful separation of the metabolite distributions into individual images having high spatial resolution.This method demonstrated the ability to provide accelerated metabolite imaging in hyperpolarized (13) C MR using multichannel coils, tailored readout, and specialized RF pulses. Magn Reson Med 74:978-989, 2015. © 2014 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.25473
View details for Web of Science ID 000364215200009
View details for PubMedID 25298086
View details for PubMedCentralID PMC4390401
-
Non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder acquisition
MAGNETIC RESONANCE IN MEDICINE
2015; 74 (3): 727-738
Abstract
To develop a new sequence for non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder (SLINCYL) acquisition.A venous saturation pulse was incorporated into a three-dimensional magnetization-prepared balanced steady-state free precession sequence for non-contrast-enhanced peripheral angiography to improve artery-vein contrast. The SLINCYL acquisition, which consists of a series of overlapped thin slabs for volumetric coverage similar to the original sliding interleaved ky (SLINKY) acquisition, was used to evenly distribute the venous-suppression effects over the field of view. In addition, the thin-slab-scan nature of SLINCYL and the centric-ordered sampling geometry of its readout trajectory were exploited to implement efficient fluid-suppression and parallel imaging schemes. The sequence was tested in healthy subjects and a patient.Compared to a multiple overlapped thin slab acquisition, both SLINKY and SLINCYL suppressed the venetian blind artifacts and provided similar artery-vein contrast. However, SLINCYL achieved this with shorter scan times and less noticeable artifacts from k-space amplitude modulation than SLINKY. The fluid-suppression and parallel imaging schemes were also validated. A patient study using the SLINCYL-based sequence well identified stenoses at the superficial femoral arteries, which were also confirmed with digital subtraction angiography.Non-contrast-enhanced angiography using SLINCYL can provide angiograms with improved artery-vein contrast in the lower extremities. Magn Reson Med 74:727-738, 2015. © 2014 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.25452
View details for Web of Science ID 000360222900014
View details for PubMedCentralID PMC4362915
-
Non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder acquisition.
Magnetic resonance in medicine
2015; 74 (3): 727-738
Abstract
To develop a new sequence for non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder (SLINCYL) acquisition.A venous saturation pulse was incorporated into a three-dimensional magnetization-prepared balanced steady-state free precession sequence for non-contrast-enhanced peripheral angiography to improve artery-vein contrast. The SLINCYL acquisition, which consists of a series of overlapped thin slabs for volumetric coverage similar to the original sliding interleaved ky (SLINKY) acquisition, was used to evenly distribute the venous-suppression effects over the field of view. In addition, the thin-slab-scan nature of SLINCYL and the centric-ordered sampling geometry of its readout trajectory were exploited to implement efficient fluid-suppression and parallel imaging schemes. The sequence was tested in healthy subjects and a patient.Compared to a multiple overlapped thin slab acquisition, both SLINKY and SLINCYL suppressed the venetian blind artifacts and provided similar artery-vein contrast. However, SLINCYL achieved this with shorter scan times and less noticeable artifacts from k-space amplitude modulation than SLINKY. The fluid-suppression and parallel imaging schemes were also validated. A patient study using the SLINCYL-based sequence well identified stenoses at the superficial femoral arteries, which were also confirmed with digital subtraction angiography.Non-contrast-enhanced angiography using SLINCYL can provide angiograms with improved artery-vein contrast in the lower extremities. Magn Reson Med 74:727-738, 2015. © 2014 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.25452
View details for PubMedID 25203505
-
Interventional Device Visualization with Toroidal Transceiver and Optically Coupled Current Sensor for Radiofrequency Safety Monitoring
MAGNETIC RESONANCE IN MEDICINE
2015; 73 (3): 1315-1327
Abstract
The development of catheters and guidewires that are safe from radiofrequency (RF) -induced heating and clearly visible against background tissue is a major challenge in interventional MRI. An interventional imaging approach using a toroidal transmit-receive (transceive) coil is presented. This toroidal transceiver allows controlled, low levels of RF current to flow in the catheter/guidewire for visualization, and can be used with conductive interventional devices that have a localized low-impedance tip contact.Toroidal transceivers were built, and phantom experiments were performed to quantify transmit power levels required for device visibility and to detect heating hazards. Imaging experiments in a pig cadaver tested the extendibility to higher field strength and nonphantom settings. A photonically powered optically coupled toroidal current sensor for monitoring induced RF currents was built, calibrated, and tested using an independent image-based current estimation method.Results indicate that high signal-to-noise ratio visualization is achievable using milliwatts of transmit power-power levels orders of magnitude lower than levels that induce measurable heating in phantom tests. Agreement between image-based current estimates and RF current sensor measurements validates sensor accuracy.The toroidal transceiver, integrated with power and current sensing, could offer a promising platform for safe and effective interventional device visualization.
View details for DOI 10.1002/mrm.25187
View details for Web of Science ID 000350279900046
View details for PubMedID 24691876
View details for PubMedCentralID PMC4182300
-
B-1 Estimation Using Adiabatic Refocusing: BEAR
MAGNETIC RESONANCE IN MEDICINE
2014; 72 (5): 1302-1310
Abstract
Accurate measurement of the nonuniform transmit radiofrequency field is useful for many applications in magnetic resonance imaging, such as calibrating the scanner transmit system, evaluating coil performance, and improving image quality and quantitation. The radiofrequency field excitation amplitude (B(1)) is often obtained by acquiring a B(1) map. In this study, a new B(1) mapping method is proposed.The use of two adiabatic full passage pulses with different magnitudes applied as successive refocusing pulses results in a linear relationship between phase and B(1) field strength that is insensitive to the repetition time, off-resonance effects, T(1), and T(2). Using this method, B(1) mapping can be localized to a slice or three-dimensional (3D) volume, with a spin-echo acquisition that is appropriate for fast projection measurements.This new method is shown to agree well with the Bloch-Siegert B(1) mapping method for both phantom and in vivo B(1) measurements at 1.5T, 3T, and 7T. The method's ability to acquire accurate projection B(1) measurements is also demonstrated.This method's high dynamic range, ability to make fast projection measurements, and linear quantitative relationship between phase and B1 make it an ideal candidate for use in robust transmitter gain calibration.
View details for DOI 10.1002/mrm.25049
View details for Web of Science ID 000343873900012
View details for PubMedCentralID PMC4031300
-
Optimization of magnetization-prepared 3-dimensional fluid attenuated inversion recovery imaging for lesion detection at 7 T.
Investigative radiology
2014; 49 (5): 290-298
Abstract
The aim of this study was to optimize the 3-dimensional (3D) fluid attenuated inversion recovery (FLAIR) pulse sequence for isotropic high-spatial-resolution imaging of white matter (WM) and cortical lesions at 7 T.We added a magnetization-prepared (MP) FLAIR module to a Cube 3D fast spin echo sequence and optimized the refocusing flip angle train using extended phase graph simulations, taking into account image contrast, specific absorption rate (SAR), and signal-to-noise ratio (SNR) as well as T1/T2 values of the different species of interest (WM, grey matter, lesions) at 7 T. We also effected improved preparation homogeneity at 7 T by redesigning the refocusing pulse used in the MP segments. Two sets of refocusing flip angle trains-(a) an SNR-optimal and (b) a contrast-optimal set-were derived and used to scan 7 patients with Alzheimer disease/cognitive impairment and 7 patients with multiple sclerosis. Conventional constant refocusing flip MP-FLAIR images were also acquired for comparison. Lesion SNR, contrast, and lesion count were compared between the 2 optimized and the standard FLAIR sequences.Whole brain coverage with 0.8 mm isotropic spatial resolution in ∼5-minute scan times was achieved using the optimized 3D FLAIR sequences at clinically acceptable SAR levels. The SNR efficiency of the SNR-optimal sequence was significantly better than that of conventional constant refocusing flip MP-FLAIR sequence, whereas the scan time was reduced more than 2-fold (∼5 vs >10 minutes). The contrast efficiency of the contrast-optimal sequence was comparable with that of the constant refocusing flip sequence. Lesion load ascertained by lesion counting was not significantly different among the sequences.Magnetization-prepared FLAIR-Cube with refocusing flip angle trains optimized for SNR and contrast can be used to characterize WM and cortical lesions at 7 T with 0.8 mm isotropic resolution in short scan times and without SAR penalty.
View details for DOI 10.1097/RLI.0000000000000041
View details for PubMedID 24566291
-
Quantitative Measurement of Cancer Metabolism Using Stimulated Echo Hyperpolarized Carbon-13 MRS
MAGNETIC RESONANCE IN MEDICINE
2014; 71 (1): 1-11
Abstract
Magnetic resonance spectroscopy of hyperpolarized substrates allows for the observation of label exchange catalyzed by enzymes providing a powerful tool to investigate tissue metabolism and potentially kinetics in vivo. However, the accuracy of current methods to calculate kinetic parameters has been limited by T1 relaxation effects, extracellular signal contributions, and reduced precision at lower signal-to-noise ratio.To address these challenges, we investigated a new modeling technique using metabolic activity decomposition-stimulated echo acquisition mode. The metabolic activity decomposition-stimulated echo acquisition mode technique separates exchanging from nonexchanging metabolites providing twice the information as conventional techniques.This allowed for accurate measurements of rates of conversion and of multiple T1 values simultaneously using a single acquisition.The additional measurement of T1 values for the reaction metabolites provides further biological information about the cellular environment of the metabolites. The new technique was investigated through simulations and in vivo studies of transgenic mouse models of cancer demonstrating improved assessments of kinetic rate constants and new T1 relaxation value measurements for hyperpolarized (13) C-pyruvate, (13) C-lactate, and (13) C-alanine.
View details for DOI 10.1002/mrm.24634
View details for Web of Science ID 000328580300001
View details for PubMedID 23412881
View details for PubMedCentralID PMC3659200
-
Optimal variable flip angle schemes for dynamic acquisition of exchanging hyperpolarized substrates
JOURNAL OF MAGNETIC RESONANCE
2013; 234: 75-81
Abstract
In metabolic MRI with hyperpolarized contrast agents, the signal levels vary over time due to T1 decay, T2 decay following RF excitations, and metabolic conversion. Efficient usage of the nonrenewable hyperpolarized magnetization requires specialized RF pulse schemes. In this work, we introduce two novel variable flip angle schemes for dynamic hyperpolarized MRI in which the flip angle is varied between excitations and between metabolites. These were optimized to distribute the magnetization relatively evenly throughout the acquisition by accounting for T1 decay, prior RF excitations, and metabolic conversion. Simulation results are presented to confirm the flip angle designs and evaluate the variability of signal dynamics across typical ranges of T1 and metabolic conversion. They were implemented using multiband spectral-spatial RF pulses to independently modulate the flip angle at various chemical shift frequencies. With these schemes we observed increased SNR of [1-(13)C]lactate generated from [1-(13)C]pyruvate, particularly at later time points. This will allow for improved characterization of tissue perfusion and metabolic profiles in dynamic hyperpolarized MRI.
View details for DOI 10.1016/j.jmr.2013.06.003
View details for Web of Science ID 000323085800009
View details for PubMedID 23845910
View details for PubMedCentralID PMC3765634
-
Adiabatic RF pulse design for Bloch-Siegert B-1(+) mapping
MAGNETIC RESONANCE IN MEDICINE
2013; 70 (3): 829-835
Abstract
The Bloch-Siegert (B-S) B 1+ mapping method has been shown to be fast and accurate, yet it suffers from high Specific Absorption Rate (SAR) and moderately long echo time. An adiabatic RF pulse design is introduced here for optimizing the off-resonant B-S RF pulse to achieve more B-S B 1+ measurement sensitivity for a given pulse width. The extra sensitivity can be used for higher angle-to-noise ratio B 1+ maps or traded off for faster scans. Using numerical simulations and phantom experiments, it is shown that a numerically optimized 2-ms adiabatic B-S pulse is 2.5 times more efficient than a conventional 6-ms Fermi-shaped B-S pulse. The adiabatic B-S pulse performance is validated in a phantom, and in vivo brain B 1+ mapping at 3T and 7T are shown. Magn Reson Med, 2012. © 2012 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.24507
View details for Web of Science ID 000323543600025
-
Perfusion and diffusion sensitive C-13 stimulated-echo MRSI for metabolic imaging of cancer
MAGNETIC RESONANCE IMAGING
2013; 31 (5): 635-642
Abstract
Metabolic imaging with hyperpolarized [1-(13)C]-pyruvate can rapidly probe tissue metabolic profiles in vivo and has been shown to provide cancer imaging biomarkers for tumor detection, progression, and response to therapy. This technique uses a bolus injection followed by imaging within 1-2 minutes. The observed metabolites include vascular components and their generation is also influenced by cellular transport. These factors complicate image interpretation, especially since [1-(13)C]lactate, a metabolic product that is a biomarker of cancer, is also produced by red blood cells. It would be valuable to understand the distribution of metabolites between the vasculature, interstitial space, and intracellular compartments. The purpose of this study was to better understand this compartmentalization by using a perfusion and diffusion-sensitive stimulated-echo acquisition mode (STEAM) MRSI acquisition method tailored to hyperpolarized substrates. Our results in mouse models showed that among metabolites, the injected substrate (13)C-pyruvate had the largest vascular fraction overall while (13)C-alanine had the smallest vascular fraction. We observed a larger vascular fraction of pyruvate and lactate in the kidneys and liver when compared to back muscle and prostate tumor tissue. Our data suggests that (13)C-lactate in prostate tumor tissue voxels was the most abundant labeled metabolite intracellularly. This was shown in STEAM images that highlighted abnormal cancer cell metabolism and suppressed vascular (13)C metabolite signals.
View details for DOI 10.1016/j.mri.2012.10.020
View details for Web of Science ID 000319103000001
View details for PubMedID 23260391
View details for PubMedCentralID PMC3626756
-
A rapid method for direct detection of metabolic conversion and magnetization exchange with application to hyperpolarized substrates
JOURNAL OF MAGNETIC RESONANCE
2012; 225: 71-80
Abstract
In this work, we present a new MR spectroscopy approach for directly observing nuclear spins that undergo exchange, metabolic conversion, or, generally, any frequency shift during a mixing time. Unlike conventional approaches to observe these processes, such as exchange spectroscopy (EXSY), this rapid approach requires only a single encoding step and thus is readily applicable to hyperpolarized MR in which the magnetization is not replenished after T(1) decay and RF excitations. This method is based on stimulated-echoes and uses phase-sensitive detection in conjunction with precisely chosen echo times in order to separate spins generated during the mixing time from those present prior to mixing. We are calling the method Metabolic Activity Decomposition Stimulated-echo Acquisition Mode or MAD-STEAM. We have validated this approach as well as applied it in vivo to normal mice and a transgenic prostate cancer mouse model for observing pyruvate-lactate conversion, which has been shown to be elevated in numerous tumor types. In this application, it provides an improved measure of cellular metabolism by separating [1-(13)C]-lactate produced in tissue by metabolic conversion from [1-(13)C]-lactate that has flowed into the tissue or is in the blood. Generally, MAD-STEAM can be applied to any system in which spins undergo a frequency shift.
View details for DOI 10.1016/j.jmr.2012.09.014
View details for Web of Science ID 000312051700011
View details for PubMedID 23143011
View details for PubMedCentralID PMC3531583
-
Reducing artifacts in one-dimensional Fourier velocity encoding for fast and pulsatile flow
MAGNETIC RESONANCE IN MEDICINE
2012; 68 (6): 1876-1885
Abstract
When evaluating the severity of valvular stenosis, the peak velocity of the blood flow is routinely used to estimate the transvalvular pressure gradient. One-dimensional Fourier velocity encoding effectively detects the peak velocity with an ungated time series of spatially resolved velocity spectra in real time. However, measurement accuracy can be degraded by the pulsatile and turbulent nature of stenotic flow and the existence of spatially varying off-resonance. In this work, we investigate the feasibility of improving the peak velocity detection capability of one-dimensional Fourier velocity encoding for stenotic flow using a novel echo-shifted interleaved readout combined with a variable-density circular k-space trajectory. The shorter echo and readout times of the echo-shifted interleaved acquisitions are designed to reduce sensitivity to off-resonance. Preliminary results from limited phantom and in vivo results also indicate that some artifacts from pulsatile flow appear to be suppressed when using this trajectory compared to conventional single-shot readouts, suggesting that peak velocity detection may be improved. The efficiency of the new trajectory improves the temporal and spatial resolutions. To realize the proposed readout, a novel multipoint-traversing algorithm is introduced for flexible and automated gradient-waveform design.
View details for DOI 10.1002/mrm.24212
View details for Web of Science ID 000311398600021
View details for PubMedID 22457248
View details for PubMedCentralID PMC3499673
-
RF pulse optimization for Bloch-Siegert B-1(+) mapping
MAGNETIC RESONANCE IN MEDICINE
2012; 68 (3): 857-862
Abstract
The Bloch-Siegert (B-S) method of B ₁⁺ mapping has been shown to be fast and accurate, yet has high SAR and moderately long TE. These limitations can lengthen scan times and incur signal loss due to B(0) inhomogeneity, particularly at high field. The B-S method relies on applying a band-limited off-resonant B-S radiofrequency pulse to induce a B ₁⁺-dependent frequency-shift for resonant spins. A method for optimizing the B-S radiofrequency pulse is presented here, which maximizes B-S B ₁⁺ measurement sensitivity for a given SAR and T(2) . A 4-ms optimized pulse is shown to have 35% less SAR compared with the conventional 6-ms Fermi pulse while still improving B ₁⁺ map angle-to-noise ratio by 22%. The optimized pulse performance is validated both in phantom and in vivo brain imaging at 7 T.
View details for DOI 10.1002/mrm.23271
View details for Web of Science ID 000308098100022
View details for PubMedID 22144397
View details for PubMedCentralID PMC3297726
-
RF Field Visualization of RF Ablation at the Larmor Frequency
IEEE TRANSACTIONS ON MEDICAL IMAGING
2012; 31 (4): 938-947
Abstract
Radio-frequency ablation (RFA) is an effective minimally invasive treatment for tumors. One primary source of difficulty is monitoring and controlling the ablation region. Currently, RFA is performed at 460 kHz, for which magnetic resonance imaging (MRI) could play a role given its capability for temperature monitoring and tumor visualization. If instead the ablation were to be performed at the MRI Larmor frequency, then the MR capability for B(1) field mapping could be used to directly visualize the radio-frequency (RF) fields created by the ablation currents. Visualizing the RF fields may enable better control of the ablation currents, enabling better control of lesion shape and size and improving repeatability. We demonstrate the feasibility of performing RFAs at 64 MHz and show preliminary results from imaging the RF fields from the ablation. The post-ablation RF fields show an increase in current density in the ablated region, consistent with an increase in conductivity of the ablated tissue.
View details for DOI 10.1109/TMI.2011.2162248
View details for Web of Science ID 000302547400008
View details for PubMedID 21775256
View details for PubMedCentralID PMC3321073
-
A method for simultaneous echo planar imaging of hyperpolarized C-13 pyruvate and C-13 lactate
JOURNAL OF MAGNETIC RESONANCE
2012; 217: 41-47
Abstract
A rapid echo planar imaging sequence for dynamic imaging of [1-(13)C] lactate and [1-(13)C] pyruvate simultaneously was developed. Frequency-based separation of these metabolites was achieved by spatial shifting in the phase-encoded direction with the appropriate choice of echo spacing. Suppression of the pyruvate-hydrate and alanine resonances is achieved through an optimized spectral-spatial RF waveform. Signal sampling efficiency as a function of pyruvate and lactate excitation angle was simulated using two site exchange models. Dynamic imaging is demonstrated in a transgenic mouse model, and phantom validations of the RF pulse frequency selectivity were performed.
View details for DOI 10.1016/j.jmr.2012.02.008
View details for Web of Science ID 000303083500007
View details for PubMedID 22405760
View details for PubMedCentralID PMC3326401
-
VERSE-guided numerical RF pulse design: A fast method for peak RF power control
MAGNETIC RESONANCE IN MEDICINE
2012; 67 (2): 353-362
Abstract
In parallel excitation, the computational speed of numerical radiofrequency (RF) pulse design methods is critical when subject dependencies and system nonidealities need to be incorporated on-the-fly. One important concern with optimization-based methods is high peak RF power exceeding hardware or safety limits. Hence, online controllability of the peak RF power is essential. Variable-rate selective excitation pulse reshaping is ideally suited to this problem due to its simplicity and low computational cost. In this work, we first improve the fidelity of variable-rate selective excitation implementation for discrete-time waveforms through waveform oversampling such that variable-rate selective excitation can be robustly applied to numerically designed RF pulses. Then, a variable-rate selective excitation-guided numerical RF pulse design is suggested as an online RF pulse design framework, aiming to simultaneously control peak RF power and compensate for off-resonance.
View details for DOI 10.1002/mrm.23010
View details for Web of Science ID 000299376500010
View details for PubMedID 22135085
View details for PubMedCentralID PMC3644517
-
Generating Super Stimulated-Echoes in MRI and Their Application to Hyperpolarized C-13 Diffusion Metabolic Imaging
IEEE TRANSACTIONS ON MEDICAL IMAGING
2012; 31 (2): 265-275
Abstract
Stimulated-echoes in MR can be used to provide high sensitivity to motion and flow, creating diffusion and perfusion weighting as well as T(1) contrast, but conventional approaches inherently suffer from a 50% signal loss. The super stimulated-echo, which uses a specialized radio-frequency (RF) pulse train, has been proposed in order to improve the signal while preserving motion and T(1) sensitivity. This paper presents a novel and straightforward method for designing the super stimulated-echo pulse train using inversion pulse design techniques. This method can also create adiabatic designs with an improved response to RF transmit field variations. The scheme was validated in phantom experiments and shown in vivo to improve signal-to-noise ratio (SNR). We have applied a super stimulated-echo to metabolic MRI with hyperpolarized (13)C-labeled molecules. For spectroscopic imaging of hyperpolarized agents, several repetition times are required but only a single stimulated-echo encoding is feasible, which can lead to unwanted motion blurring. To address this, a super stimulated-echo preparation scheme was used in which the diffusion weighting is terminated prior to the acquisition, and we observed a SNR increases of 60% in phantoms and 49% in vivo over a conventional stimulated-echo. Experiments following injection of hyperpolarized [1-(13)C] -pyruvate in murine transgenic cancer models have shown improved delineation for tumors since signals from metabolites within tumor tissues are retained while those from the vasculature are suppressed by the diffusion preparation scheme.
View details for DOI 10.1109/TMI.2011.2168235
View details for Web of Science ID 000300197500010
View details for PubMedID 22027366
View details for PubMedCentralID PMC3274664
-
Fast Dynamic 3D MR Spectroscopic Imaging With Compressed Sensing and Multiband Excitation Pulses for Hyperpolarized C-13 Studies
MAGNETIC RESONANCE IN MEDICINE
2011; 65 (3): 610-619
Abstract
Hyperpolarized 13C MR spectroscopic imaging can detect not only the uptake of the pre-polarized molecule but also its metabolic products in vivo, thus providing a powerful new method to study cellular metabolism. Imaging the dynamic perfusion and conversion of these metabolites provides additional tissue information but requires methods for efficient hyperpolarization usage and rapid acquisitions. In this work, we have developed a time-resolved 3D MR spectroscopic imaging method for acquiring hyperpolarized 13C data by combining compressed sensing methods for acceleration and multiband excitation pulses to efficiently use the magnetization. This method achieved a 2 sec temporal resolution with full volumetric coverage of a mouse, and metabolites were observed for up to 60 sec following injection of hyperpolarized [1-(13)C]-pyruvate. The compressed sensing acquisition used random phase encode gradient blips to create a novel random undersampling pattern tailored to dynamic MR spectroscopic imaging with sampling incoherency in four (time, frequency, and two spatial) dimensions. The reconstruction was also tailored to dynamic MR spectroscopic imaging by applying a temporal wavelet sparsifying transform to exploit the inherent temporal sparsity. Customized multiband excitation pulses were designed with a lower flip angle for the [1-(13)C]-pyruvate substrate given its higher concentration than its metabolic products ([1-(13)C]-lactate and [1-(13)C]-alanine), thus using less hyperpolarization per excitation. This approach has enabled the monitoring of perfusion and uptake of the pyruvate, and the conversion dynamics to lactate and alanine throughout a volume with high spatial and temporal resolution.
View details for DOI 10.1002/mrm.22650
View details for Web of Science ID 000287929800002
View details for PubMedID 20939089
View details for PubMedCentralID PMC3021589
-
Frequency-Offset Cartesian Feedback for MRI Power Amplifier Linearization
IEEE TRANSACTIONS ON MEDICAL IMAGING
2011; 30 (2): 512-522
Abstract
High-quality magnetic resonance imaging (MRI) requires precise control of the transmit radio-frequency (RF) field. In parallel excitation applications such as transmit SENSE, high RF power linearity is essential to cancel aliased excitations. In widely-employed class AB power amplifiers, gain compression, cross-over distortion, memory effects, and thermal drift all distort the RF field modulation and can degrade image quality. Cartesian feedback (CF) linearization can mitigate these effects in MRI, if the quadrature mismatch and dc offset imperfections inherent in the architecture can be minimized. In this paper, we present a modified Cartesian feedback technique called "frequency-offset Cartesian feedback" (FOCF) that significantly reduces these problems. In the FOCF architecture, the feedback control is performed at a low intermediate frequency rather than dc, so that quadrature ghosts and dc errors are shifted outside the control bandwidth. FOCF linearization is demonstrated with a variety of typical MRI pulses. Simulation of the magnetization obtained with the Bloch equation demonstrates that high-fidelity RF reproduction can be obtained even with inexpensive class AB amplifiers. Finally, the enhanced RF fidelity of FOCF over CF is demonstrated with actual images obtained in a 1.5 T MRI system.
View details for DOI 10.1109/TMI.2010.2087768
View details for Web of Science ID 000286931000029
View details for PubMedID 20959264
View details for PubMedCentralID PMC3155726
-
Minimum Envelope Roughness Pulse Design for Reduced Amplifier Distortion in Parallel Excitation
MAGNETIC RESONANCE IN MEDICINE
2010; 64 (5): 1433-1440
Abstract
Parallel excitation uses multiple transmit channels and coils, each driven by independent waveforms, to afford the pulse designer an additional spatial encoding mechanism that complements gradient encoding. In contrast to parallel reception, parallel excitation requires individual power amplifiers for each transmit channel, which can be cost prohibitive. Several groups have explored the use of low-cost power amplifiers for parallel excitation; however, such amplifiers commonly exhibit nonlinear memory effects that distort radio frequency pulses. This is especially true for pulses with rapidly varying envelopes, which are common in parallel excitation. To overcome this problem, we introduce a technique for parallel excitation pulse design that yields pulses with smoother envelopes. We demonstrate experimentally that pulses designed with the new technique suffer less amplifier distortion than unregularized pulses and pulses designed with conventional regularization.
View details for DOI 10.1002/mrm.22512
View details for Web of Science ID 000283616900023
View details for PubMedCentralID PMC3053148
-
Minimum envelope roughness pulse design for reduced amplifier distortion in parallel excitation.
Magnetic resonance in medicine
2010; 64 (5): 1432-1439
Abstract
Parallel excitation uses multiple transmit channels and coils, each driven by independent waveforms, to afford the pulse designer an additional spatial encoding mechanism that complements gradient encoding. In contrast to parallel reception, parallel excitation requires individual power amplifiers for each transmit channel, which can be cost prohibitive. Several groups have explored the use of low-cost power amplifiers for parallel excitation; however, such amplifiers commonly exhibit nonlinear memory effects that distort radio frequency pulses. This is especially true for pulses with rapidly varying envelopes, which are common in parallel excitation. To overcome this problem, we introduce a technique for parallel excitation pulse design that yields pulses with smoother envelopes. We demonstrate experimentally that pulses designed with the new technique suffer less amplifier distortion than unregularized pulses and pulses designed with conventional regularization.
View details for DOI 10.1002/mrm.22512
View details for PubMedID 20632401
View details for PubMedCentralID PMC3053148
-
Investigation of Tumor Hyperpolarized [1-C-13]-Pyruvate Dynamics Using Time-Resolved Multiband RF Excitation Echo-Planar MRSI
MAGNETIC RESONANCE IN MEDICINE
2010; 63 (3): 582-591
Abstract
Hyperpolarized [1-(13)C]-pyruvate is an exciting new agent for the in vivo study of cellular metabolism and a potential cancer biomarker. The nature of the hyperpolarized signal poses unique challenges because of its short duration and the loss of magnetization with every excitation. In this study, we applied a novel and efficient time-resolved MR spectroscopic imaging (MRSI) method to investigate in a prostate cancer model the localized temporal dynamics of the uptake of [1-(13)C]-pyruvate and its conversion to metabolic products, specifically [1-(13)C]-lactate. This hyperpolarized (13)C method used multiband excitation pulses for efficient use of the magnetization. This study demonstrated that regions of tumor were differentially characterized from normal tissue by the lactate dynamics, where tumors showed later lactate detection and longer lactate duration that was statistically significant (P < 0.001). Compared to late-pathologic-stage tumors, early- to intermediate-stage tumors demonstrated significantly (P < 0.01) lower lactate total signal-to-noise ratio (SNR), with similar temporal dynamic parameters. Hyperpolarized pyruvate dynamics provided uptake, perfusion, and vascularization information on tumors and normal tissue. Large, heterogeneous tumors demonstrated spatially variable uptake of pyruvate and metabolic conversion that was consistent with cellularity and necrosis identified by histology. The results of this study demonstrated the potential of this new hyperpolarized MR dynamic method for improved cancer detection and characterization.
View details for DOI 10.1002/mrm.22264
View details for Web of Science ID 000274938000006
View details for PubMedID 20187172
View details for PubMedCentralID PMC2844437
-
Maximum Linear-Phase Spectral-Spatial Radiofrequency Pulses for Fat-Suppressed Proton Resonance Frequency-Shift MR Thermometry
MAGNETIC RESONANCE IN MEDICINE
2009; 62 (5): 1242-1250
Abstract
Conventional spectral-spatial pulses used for water-selective excitation in proton resonance frequency-shift MR thermometry require increased sequence length compared to shorter wideband pulses. This is because spectral-spatial pulses are longer than wideband pulses, and the echo time period starts midway through them. Therefore, for a fixed echo time, one must increase sequence length to accommodate conventional spectral-spatial pulses in proton resonance frequency-shift thermometry. We introduce improved water-selective spectral-spatial pulses for which the echo time period starts near the beginning of excitation. Instead of requiring increased sequence length, these pulses extend into the long echo time periods common to PRF sequences. The new pulses therefore alleviate the traditional tradeoff between sequence length and fat suppression. We experimentally demonstrate an 11% improvement in frame rate in a proton resonance frequency imaging sequence compared to conventional spectral-spatial excitation. We also introduce a novel spectral-spatial pulse design technique that is a hybrid of previous model- and filter-based techniques and that inherits advantages from both. We experimentally validate the pulses' performance in suppressing lipid signal and in reducing sequence length compared to conventional spectral-spatial pulses.
View details for DOI 10.1002/mrm.22118
View details for Web of Science ID 000271431200018
View details for PubMedID 19780177
View details for PubMedCentralID PMC2795148
-
Fast Large-Tip-Angle Multidimensional and Parallel RF Pulse Design in MRI
IEEE TRANSACTIONS ON MEDICAL IMAGING
2009; 28 (10): 1548-1559
Abstract
Large-tip-angle multidimensional radio-frequency (RF) pulse design is a difficult problem, due to the nonlinear response of magnetization to applied RF at large tip-angles. In parallel excitation, multidimensional RF pulse design is further complicated by the possibility for transmit field patterns to change between subjects, requiring pulses to be designed rapidly while a subject lies in the scanner. To accelerate pulse design, we introduce a fast version of the optimal control method for large-tip-angle parallel excitation. The new method is based on a novel approach to analytically linearizing the Bloch equation about a large-tip-angle RF pulse, which results in an approximate linear model for the perturbations created by adding a small-tip-angle pulse to a large-tip-angle pulse. The linear model can be evaluated rapidly using nonuniform fast Fourier transforms, and we apply it iteratively to produce a sequence of pulse updates that improve excitation accuracy. We achieve drastic reductions in design time and memory requirements compared to conventional optimal control, while producing pulses of similar accuracy. The new method can also compensate for nonidealities such as main field inhomogeneties.
View details for DOI 10.1109/TMI.2009.2020064
View details for Web of Science ID 000270226100004
View details for PubMedID 19447704
View details for PubMedCentralID PMC2763429
-
Spiral Imaging Artifact Reduction: A Comparison of Two k-Trajectory Measurement Methods
JOURNAL OF MAGNETIC RESONANCE IMAGING
2009; 29 (6): 1485-1492
Abstract
To compare an external sensor-based k-space calibration technique with a routine precalibration method for quantification of method accuracy and reduction of spiral imaging artifacts to obtain improved image quality.Recently, magnetic field monitoring (MFM) has been introduced as a new calibration technique of gradient field-related imperfections. External sensors are placed near the observed object to measure magnetic field variations during image acquisition. The measured field data are used to determine the actual k-space trajectory and for image reconstruction to reduce artifacts. In the past, precalibration techniques have been proposed where the k-space trajectory is measured by means of special calibration sequences directly in the object of interest. In this study, MFM is introduced as an effective correction technique for spiral imaging. On the basis of a comparison, whether MFM is as viable as the chosen reference method presented by Duyn et al (J Magn Reson [1998] 132:150-153) is analyzed in terms of detecting imperfections of spatial encoding gradients in order to correct for these in image reconstruction. As this technique is used as a reference method, it is given the acronym Duyn calibration technique (DCT). MFM and DCT are compared and timing delays, k-space offsets, eddy current effects, k-trajectory error propagation, image distortions, and signal-to-noise ratio were determined for different spiral sequences in two different phantoms.Both techniques effectively detect k-space offsets and k-trajectory error propagation and correct for general error sources like timing delays. In object border areas, artifacts such as deformations and blurring were dramatically reduced. Within all tested categories, MFM performed as well as DCT. In terms of k-trajectory error propagation and image distortion quantification, MFM was more accurate.We introduce MFM as an effective and accurate correction technique for spiral imaging, where a comparison of MFM and DCT has shown that both techniques are accurate correction techniques for spiral imaging.
View details for DOI 10.1002/jmri.21782
View details for Web of Science ID 000266673200032
View details for PubMedID 19472426
-
Design of Cosine Modulated Very Selective Suppression Pulses for MR Spectroscopic Imaging at 3T
MAGNETIC RESONANCE IN MEDICINE
2009; 61 (3): 533-540
Abstract
The advantages of using a 3 Tesla (T) scanner for MR spectroscopic imaging (MRSI) of brain tissue include improved spectral resolution and increased sensitivity. Very selective saturation (VSS) pulses are important for maximizing selectivity for PRESS MRSI and minimizing chemical shift misregistration by saturating signals from outside the selected region. Although three-dimensional (3D) PRESS MRSI is able to provide excellent quality metabolic data for patients with brain tumors and has been shown to be important for defining tumor burden, the method is currently limited by how much of the anatomic lesion can be covered within a single examination. In this study we designed and implemented cosine modulated VSS pulses that were optimized for 3T MRSI acquisitions. This provided improved coverage and suppression of unwanted lipid signals with a smaller number of pulses. The use of the improved pulse sequence was validated in volunteer studies, and in clinical 3D MRSI exams of brain tumors.
View details for DOI 10.1002/mrm.21842
View details for Web of Science ID 000263608300005
View details for PubMedID 19097232
View details for PubMedCentralID PMC2690719
-
Compressed sensing for resolution enhancement of hyperpolarized C-13 flyback 3D-MRSI
JOURNAL OF MAGNETIC RESONANCE
2008; 192 (2): 258-264
Abstract
High polarization of nuclear spins in liquid state through dynamic nuclear polarization has enabled the direct monitoring of 13C metabolites in vivo at very high signal-to-noise, allowing for rapid assessment of tissue metabolism. The abundant SNR afforded by this hyperpolarization technique makes high-resolution 13C 3D-MRSI feasible. However, the number of phase encodes that can be fit into the short acquisition time for hyperpolarized imaging limits spatial coverage and resolution. To take advantage of the high SNR available from hyperpolarization, we have applied compressed sensing to achieve a factor of 2 enhancement in spatial resolution without increasing acquisition time or decreasing coverage. In this paper, the design and testing of compressed sensing suited for a flyback 13C 3D-MRSI sequence are presented. The key to this design was the undersampling of spectral k-space using a novel blipped scheme, thus taking advantage of the considerable sparsity in typical hyperpolarized 13C spectra. Phantom tests validated the accuracy of the compressed sensing approach and initial mouse experiments demonstrated in vivo feasibility.
View details for DOI 10.1016/j.jmr.2008.03.003
View details for Web of Science ID 000256538300011
View details for PubMedID 18367420
View details for PubMedCentralID PMC2475338
-
In vivo real-time intravascular MRI
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE
2002; 4 (2): 223-232
Abstract
The Magnetic resonance imaging (MRI) is an emerging technology for catheter-based imaging and interventions. Real-time MRI is a promising methodfor overcoming catheter and physiologic motion for intravascular imaging.All imaging was performed on a 1.5 T Signa MRI scanner with high-speed gradients. Multiple catheter coils were designed and constructed, including low-profile, stub-matched coils. Coil sensitivity patterns and SNR measurements were compared. Real-time imaging was performed with an interleaved spiral sequence using a dedicated workstation, providing real-time data acquisition, image reconstruction and interactive control and display. Real-time "black-blood" imaging was achieved through incorporation of off-slice saturation pulses. The imaging sequence was tested in a continuous flow phantom and then in vivo in the rabbit aorta using a 2 mm catheter coil.The real-time intravascular imaging sequence achieved 120-440 micron resolution at up to 16 frames per second. Low-profile stub-tuned catheter coils achieved similar SNR to larger traditional coil designs. In the phantom experiments, addition of real-time black-blood saturation pulses effectively suppressed the flow signal and allowed visualization of the phantom wall. In vivo experiments clearly showed real-time intravascular imaging of the rabbit aortic wall with minimal motion artifacts and effective blood signal suppression.Real-time imaging with low-profile coil designs provides significant enhancements to intravascular MRI.
View details for PubMedID 12074137
-
Real-time interactive MRI on a conventional scanner
MAGNETIC RESONANCE IN MEDICINE
1997; 38 (3): 355-367
Abstract
A real-time interactive MRI system capable of localizing coronary arteries and imaging arrhythmic hearts in real-time is described. Non-2DFT acquisition strategies such as spiral-interleaf, spiral-ring, and circular echo-planar imaging provide short scan times on a conventional scanner. Real-time gridding reconstruction at 8-20 images/s is achieved by distributing the reconstruction on general-purpose UNIX workstations. An X-windows application provides interactive control. A six-interleaf spiral sequence is used for cardiac imaging and can acquire six images/s. A sliding window reconstruction achieves display rates of 16-20 images/s. This allows cardiac images to be acquired in real-time, with minimal motion and flow artifacts, and without breath holding or cardiac gating. Abdominal images are acquired at over 2.5 images/s with spiral-ring or circular echo-planar sequences. Reconstruction rates are 8-10 images/s. Rapid localization in the abdomen is demonstrated with the spiral-ring acquisition, whereas peristaltic motion in the small bowel is well visualized using the circular echo-planar sequence.
View details for Web of Science ID A1997XW16200002
View details for PubMedID 9339436